Advertisement

3)
Check the blue rubber grommets (P/N 10630) and grounding clips (P/N 10752) on the
oxygen blender mounting (4 each) for wear and replace if necessary (LTV
®
4)
For the LTV
1200: Replace the four (4) blender mounting screws (P/N 10176), O
blender cap (P/N 15016-001 or 13301-001), and four (4) flat metal washers (P/N
10594). The grounding clips should be positioned over the inside lip of the grommets,
aligned with the edge of the side panel and oriented as shown in the illustrations.
Screws should be torqued to 60 in-oz (0.42 Nm).
®
For the LTV
1150 and 1100: Replace the four (4) O
10627), Green O
10630). Screws should be torqued to 60 in-oz (0.42 Nm).
NOTE
If the upper clips are not properly aligned with the edge of the side panel, the back panel of
the ventilator will not seat correctly. Use a straight edged screwdriver or pick to prevent each
grounding clip from turning as the mounting screws are tightened (LTV
Page 8-100
port cap (P/N 14446), and four (4) plastic shoulder washers (P/N
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
Block mounting screws (P/N
2
®
1200 only).
p/n 18603-001, Rev. J
1200 only).
2