Storage and Operating Conditions (cont.)
Specification
Oxygen Inlet
DISS or NIST Connector Inlet Pressure Range
®
(LTV
1200 Only)
Tapered Tubing Connector Inlet Pressure Range
Shock and Vibration
The ventilator is designed to withstand shock and vibration in accordance with relevant
requirements set forth in the following standards:
IEC 68-2-27
IEC 68-2-6
IEC 68-2-34
MIL-STD-810E
Spillage
The ventilator resists fluid spillage when tested in accordance with the relevant standards
specified in IEC 601-1 Clause 44.3.
External Surface Temperature
External surfaces
Communications
Port
Communications
Patient Assist Call /
Remote Alarm
Equipment Classification
Classification
Type
Page A-8
Shock
Vibration
Vibration
Shock, Ground Transport and Helicopter Transport Vibration
< 50C, ambient temperature of 35C
Connector
RS232, DB9 connector
RJ11-4
The ventilator is rated as Class II equipment per IEC 601-1 Clause 6.11
The ventilator is specified as Type BF equipment per IEC 601-1 Clause 6.11
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tolerance
40 to 80 PSIG
 2 PSIG
0 to 10 PSIG
 2 PSIG
n/a
Specification
Protocol Options: Data, Monitor, Printer,
Modem
Closed contact resistance:  1 ohm
p/n 18603-001, Rev. J