12) Connect the bypass tubing from the flow valve to the barbed elbow at the bottom of the
turbine manifold. The tubing should be looped into the space between the bottom of the
manifold and the bottom edge of the upper weldment so that it is out of the way and will
not be pinched when the back of the ventilator is replaced.
13) Replace the two (2) 7/16" pan-head screws securing the accumulator to the turbine
manifold spacer. See instructions on page 8-53 Torque both screws to 60 in-oz (0.42
Nm).
14) Reconnect the 3-wire flow valve connector to the power board. The connector is keyed
to fit in only one direction and will snap into place when properly connected.
®
15) (LTV
1200 only) Insert the O
from the blender into (at least 1/2")
the Pisco connector attached to the
oxygen pressure transducer on the
analog board. Gently pull (~2-lbs
force) the tube while holding the
Pisco connector to ensure it is
properly attached.
If the tube is not inserted
correctly, it will detach when a
high pressure oxygen source
is connected to the ventilator
p/n 18603-001, Rev. J
tube
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-137