Event Codes
This section includes a list of the event codes that can be recorded in the Event Trace.
Event Codes by Code #
Code
Event Name
01
VENT 1
02
VENT 0
03
HOUR MTR
04
VENT CHK
05
APNEA 1
06
APNEA 0
07
N/A
08
HIGH DIS
09
LOW DIS
10
DISC 0
11
BATMPT1
12
BATMPT0
13
BATLOW1
14
BATLOW0
15
EXT LST1
16
EXT LST0
17
EXT LOW1
18
EXT LOW0
19
XDC FLT1
20
XDC FLT0
21
O2 LOW 1
22
O2 LOW 0
23
O2 HI 1
24
O2 HI 0
25
DEFAULTS
26
NO CAL
27
FAN FLT1
28
FAN FLT0
29
N/A
30
N/A
31
INTRRPT1
32
INTRRPT2
33
AD MMTCH
p/n 18603-001, Rev. J
Event
Power on
Power off
Set hour meter
Set vent check
Apnea mode entered
Apnea mode exited
Not used
High side disconnect
Low side disconnect
Circuit disconnect exited
Internal battery empty occurred
Internal battery empty exited
Internal battery low occurred
Internal battery low exited
External power lost occurred
External power lost exited
External power low occurred
External power low exited
XDCR fault occurred
XDCR fault exited
O
pressure low occurred
2
O
pressure low exited
2
O
pressure high occurred
2
O
pressure high exited
2
Defaults, or Set Defaults occurred
No calibration data found
Fan fault occurred
Fan fault exited
Not used
Not used
Spurious interrupt occurred ms
Spurious interrupt occurred ls
ADC mismatch
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Associated Alarm
None
None
None
Entered VENT CHECK mode
APNEA
APNEA
DISC/SENSE
DISC/SENSE
DISC/SENSE
BAT EMPTY
BAT EMPTY
BAT LOW
BAT LOW
POWER LOST
POWER LOST
POWER LOW
POWER LOW
XDCR FAULT
XDCR FAULT
LOW O2 PRES
LOW O2 PRES
HIGH O2 PRES
HIGH O2 PRES
DEFAULTS / DEFAULTS, SET
NO CAL DATA
HW FAULT
HW FAULT
RESET or RESET 1
RESET or RESET 1
HW FAULT
Page D-3