Advertisement

3)
Check the values displayed in the Tidal Volume window. When the value stabilizes,
record the number on the Calibration Worksheet and press Select.
O2 0 PSI is displayed.
4)
With the O
inlet port open to ambient room air, observe the values in the Tidal Volume
2
window. When value stabilizes, record the number on the Calibration Worksheet. Press
Select.
MOTOR DRIVE is displayed if the calibration was successful.
If CAL FAIL appears at any time, press Select to exit the calibration. When O2 INLET is
displayed, press Select and perform the calibration again.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 6-19