Solenoid Manifold Assembly

Parts required for replacement:
Solenoid Manifold Assembly P/N 18528-001

Solenoid Manifold Tube Routing Table

Solenoid
Manifold
Port Number
#1
Onto the high pressure port on the Valve Differential Transducer (PT7)
#2
Onto the Flow Differential Transducer (PT4)
#3
Onto the Airway Pressure Transducer (PT6)
#4
Onto the side panel low pressure Flow Transducer port
#5
From the Flow Valve (clear tubing)
#6
Onto side panel Exhalation Valve port
#7
Onto the Flow Differential Transducer
#8
Onto the side panel high pressure Flow Transducer port
#9
Onto Accumulator port A
NOTE
Tubing Configurations - When disconnecting tubes, note their positions and review the
internal flexible tube routing tables, instructions and diagrams beginning on page 8-33.
Page 8-120
Port Tubing Routed
through a tee & elbow subassembly
male Luer fitting on outside of panel
barbed fitting on outside of panel
through a tube & elbow subassembly
female Luer fitting on outside of panel
through a tube & tee subassembly
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with torque meter
1/4" Nut Driver adapter for torque wrench
Grounded anti-static wrist strap
Torque screwdriver (120 in-oz/0.84 Nm)
Tubing
P/N
18624-001
18337-003
18337-006
18339-003
10455
18339-005
18625-001
and
18338-001
18339-005
18626-001
p/n 18603-001, Rev. J