5) Install the current sounder assembly in the cut-out in the right side of the upper
weldment.
6) Line the sounder assembly up with the mounting hole in the side panel as shown.
7) Replace the pan-head screw in the sounder bracket and torque tighten to 60 in-oz (0.42
Nm).
8) Connect the 2-wire lead connector on the sounder assembly to the 2 pin male connector
on the power board. The small legs on the sounder assembly wire connector should be
indexed towards the vertical tab on the power board connector. When properly seated, it
will snap into place.
9) Install an O
Sampling Tube
2
Component Removal and Replacement for detailed instructions.
10) Verify that no tubing is kinked or twisted, and that it will not be pinched when
compressed by the back panel. Refer to the configuration instructions, tables and
diagrams beginning on page 8-33 to check the tubing routing. Always check ALL tube
routing for kinks or compression that might result in restricted flows prior to closing the
ventilator.
11) Reconnect the internal battery and replace the back panel (see page 8-29).
49
O
Sampling Tube, P/N 10544, ~10.0" long, 0.125" O.D. X 0.079 I.D. clear polycarbonate tubing.
2
Page 8-58
49
. See Back Panel, Reinstallation in Chapter 8 –
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J