Transducer Calibration, Acceptable A/D Counts

Airway Pressure (AP)
@ 50 cmH
O
2
Ambient
Flow Differential Bi-Directional (FDb)
@ -30 cmH
O
2
Ambient
Flow Differential Wide (FDw)
@ 30 cmH
O
2
Ambient
Flow Differential Narrow (FDn)
80 ± 70
Ambient
4015 ± 70
Ambient
Flow Valve Differential (FVd)
@ 15 cmH
O
2
Ambient (LTV 1200/1150)
Ambient (LTV 1100)
PEEP Pilot Pressure (PPP)
@ 30 cmH
O (LTV 1200/1150)
2
@ 30 cmH
O (LTV 1100)
2
Ambient (LTV 1200/1150)
Ambient (LTV 1100)
Oxygen Pressure (O
) (LTV
2
@ 50 PSI
Ambient
Page E-16
733 – 1570
35-350
64-3240
3180-4045
854 – 4030
10 – 400
10 - 150
128-3968
3945 - 4085
128-3968
1915 – 2340 (For Software Version < 05.04)
1915 – 2521 (For Software Version ≥ 05.04)
40 – 328
61 – 354
999 – 2493
1039 – 3082
135 – 272
133 – 280
®
1200 only)
900 – 1822
122 – 246
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J