Vent Inop Alarm Test
The Vent Inop alarm test is used to verify that the Vent Inop alarm is working correctly.
To run the Vent Inop alarm test:
1)
To run the Vent Inop alarm test, the ventilator must be on (running) for at least 60
seconds.
2)
Turn the ventilator off by pressing and holding On/Standby for a minimum of 3
seconds. DO NOT press Silence Reset.
3)
Observe the ventilator for 15 seconds and verify that both of the following occur;
The alarm tone sounds continuously for a full 15 seconds.
The Vent Inop LED illuminates continuously for a full 15 seconds.
4)
When the test is complete, silence the alarm by pressing Silence Reset.
5)
Verify a confirming audible chirp sounds after the alarm is silenced.
If the ventilator fails the Vent Inop alarm test, see Chapter 7 - Troubleshooting for more
information.
NOTE
®
Although the LTV
1200 Front Panel is shown, the test is applicable to the LTV
1100 Ventilators.
Page 2-10
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
1150 and
p/n 18603-001, Rev. J