18) Position the back surface of the lower portion of the tool against the turbine manifold
with the lowest edge of the tool resting on the main PCBA below the power board. Hold
the lower portion of the tool against the turbine manifold and rotate the handle
backwards and down toward the top surface of the turbine manifold to separate the
power and main PCBA mating connectors (see Cut-away illustration below).
19) Lift out the power board, flexing the left side panel slightly out of the way.