To permanently remove the lower protective boot
1)
Lay the ventilator down (front up) and use a Philips-head screwdriver to remove the four
flat-head mounting screws and finish washers in the sides of the Lower Protective Boot,
as indicated in the illustration below.
2)
Remove the lower boot and insert and thread four 1/4" flat-head mounting screws with
finish washers into the screw holes in the ventilator's side panels, as indicated in the
illustration below.
Finish washers should be already in place.
3)
Torque tighten all four screws to 20 in-oz (0.14 Nm)
CAUTION
Damage to finish washers -Do not over tighten screws to avoid damage to the finish
washers.
33
See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for
additional information.