CareFusion LTV 1200 Service Manual page 197

Table of Contents

Advertisement

10) Remove the accumulator from the turbine manifold spacer.
11) Remove the four (4) screws securing the turbine manifold spacer to the turbine manifold
assembly. Loosen or remove the single screw securing the solenoid mount assembly
and remove the two screws holding down the accumulator. Move the solenoid mount
assembly and the accumulator from the spacer. Remove the spacer and set aside away
from the ventilator.
NOTE
The analog board, solenoid manifold and solenoid mount assembly cable may be left
attached to the power board.
12) Remove the power board (see instructions on page 8-105). The analog board may be
left attached to the power board. Be careful not to catch the power board on the
sounder bracket while removing it.
Page 8-76
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J