Settings, Front Panel Controls And Extended Features - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Settings, Front Panel Controls and Extended Features

The following tables provide factory default information.
Control - Default
Breath Rate - 12 bpm
Control Lock - On
Data Display Scrolling - Auto-On
High Pressure Limit - 20 cmH
Inspiratory/Expiratory Hold - Off
Inspiratory Time - 1.5 sec
Low Minute Volume - 2.5 Lpm
Low Pressure - 5 cmH
Low Pressure O
Source - Off
2
The factory-set default Extended Features settings are:
Feature - Default
Alarm Volume - 85 dBA
Apnea Interval - 20 sec
106
Com Setting
- Data
Control Unlock - Easy
Date Format - MM/DD/YYYY
High f Alarm - High f Off
High f Alarm Delay - 30 sec
High PEEP Alarm - PEEP +5 cmH
HP Alarm Delay - No Delay
Language - English
Leak Compensation - On
Leak Test Query - Off
Low PEEP Alarm - PEEP -5 cmH
LPP Alarm - All Breaths
NPPV Mode - Off
O
Conserve - Off
2
O
Duration Cylinder - 2000 psi or 138 bar
2
O
Duration Cylinder - 622 liters
2
O
Flush Period - 3 min
2
Patient Assist - Normal
106
This feature is not reset to default values when the SET DEFAULTS option is used in Extended Features.
p/n 18603-001, Rev. J
O
2
O
2
O
2
O
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Control - Default
O
% (O
Flush) - 21%
2
2
PEEP - 0 cmH
Pressure Control - 1 cmH
Pressure Support - 1 cmH
Sensitivity - 2 Lpm
Tidal Volume - 500 ml
Ventilation Mode - Assist/Control
Volume Pressure Mode - Volume
Feature - Default
Patient Circuit Mode - Flow Sensor
Patient Circuit Query - Off
Patient Query - On
Patient Size - Infant
PC Flow Termination - Off
PIP LED - On
Rise Time Profile - 4
SBT Display f/Vt - On
SBT FIO
SBT High f Alarm - 35 bpm
SBT High f/Vt Alarm - 105 bpm / l
SBT Low f Alarm - 10 bpm
SBT Low f/Vt Alarm - 70 bpm / l
SBT Mode - Off
SBT Mode Run Time - 20 min
SBT PEEP - 0 cmH
SBT Pressure Support - 10 cmH
Sigh Breath - Off
Variable Flow Termination - 25%
Variable Time Termination - 1.5 sec
O
2
O
2
O
2
- 21%
2
O
2
O
2
Page E-11