Thermo Conductive Turbine Pad

Thermo conductive pads should be replaced any time they have hardened, and at least at
the 30,000 hour service. The turbine heat sink pad is located between the inside of the
upper weldment and the turbine.
Parts required for replacement:
Thermo Conductive Turbine Pad P/N
10129
To replace the thermo conductive turbine pad, see instructions for removing and replacing
the turbine manifold on page 8-131.
Page 8-130
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with torque meter
Grounded anti-static wrist strap
Mild cleanser
p/n 18603-001, Rev. J