6)
Calculate the T/S value measured in revolutions per minute (RPM) by multiplying the
frequency displayed on the meter by the number 30. Round the total to the nearest 10.
7)
If the measured T/S value is less than 4000, then subtract it from 4000 and add this
adjustment value to 4000.
If the measured T/S value is greater than 4000, then subtract 4000 and subtract this
adjustment value from 4000.
Example:
If the measured T/S value is 3750, then the actual turbine speed
setting should be 4250.
If the measured T/S value is 4250, then the actual turbine speed
setting should be 3750.
8)
The actual T/S shall be 3600 to 4600 RPM.
9)
Record the actual T/S value on the Calibration Worksheet. This value is used in the
Performance Checkout patient port pressure testing.
If the actual T/S value is outside of the above range, check the setup and repeat the test.
For assistance, call Technical Support at CareFusion using the contact information at the
front of this manual.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 6-23