30) Ensure that the flexible tube from port B of the accumulator is tucked between the purge
solenoids and the turbine manifold. Replace the turbine manifold spacer using four (4)
7/16" pan-head screws (P/N 10433), Torque-tightened to 60 in-oz (0.42 Nm).
31) Replace the accumulator (see Accumulator on page 8-53).
32) Replace the solenoid mount assembly (see Solenoid Mount Assembly on page 8-126).
33) Replace the motor board. Use care when aligning the pass through connectors (see
instructions on page 8-94).
34) Reinstall the flow valve assembly and reconnect the orange and clear flow valve tubing
(see instructions on page 8-66).
35) Reconnect the 3-wire flow valve connector to the power board. The connector can only
be installed in one direction and will snap into place when properly connected.
36) Reconnect the Pisco connector to the oxygen pressure transducer on the analog board
®
(LTV
1200 only).
37) Reconnect the 4-wire connector from the flow valve and the 3-wire and 5-wire
connectors from the turbine to the motor board. These connectors can only be installed
in one direction and will snap into place when properly connected.
38) Reconnect the 2-wire flow valve temperature connector to the power board. The
connector can only be installed in one direction and will snap into place when properly
connected.
39) Install an O
Sampling Tube
2
Component Removal and Replacement for detailed instructions.
40) Verify that no tubing is kinked or twisted, and that it will not be pinched when
compressed by the back panel. Refer to the routing instructions, tables and diagrams
beginning on page 8-33 to check the tubing routing. Always check ALL tube routing for
kinks or compression that might result in restricted flows prior to closing the ventilator.
41) Reconnect the internal battery and replace the back panel (see page 8-29).
58
O
Sampling Tube, P/N 10544, ~10.0" long, 0.125" O.D. X 0.079 I.D. clear polycarbonate tubing.
2
p/n 18603-001, Rev. J
58
. See Back Panel, Reinstallation in Chapter 8 –
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-91