Settings, Dip Switch
The following Dip Switch Settings table provides DIP Switch factory set default information.
Dip Switch
No.
1
Force all LEDs to turn on. Use in conjunction with dip switch #5.
2
Force O
3
Disallow dimming when on battery and no Front Panel activity after
60 seconds.
4
5
OFF = Normal operation.
ON = Maintenance mode (for calibration)
6
RTC clock battery connection.
7
ON = Flash write enabled.
8
ON = Flash write protect.
Page E-10
Intended for factory use only
pressure to 50 psi regardless of inlet pressure.
2
Intended for factory use only
Intended for factory use only
Intended for factory use only
Intended for factory use only
Intended for factory use only
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Function
Default
Setting
OFF
OFF
OFF
OFF
OFF
ON
OFF
ON
p/n 18603-001, Rev. J