Back Panel, Reinstallation

Parts required for replacement:
Replace if damaged:
1/4" Flat head Screw (3) P/N 10430
1/2" Flat head Screw (6) P/N 10338
Finish Washer (6) P/N 10191, or 19119-002
Grounding Clips (3) P/N 10752
Thermo Conductive Pad P/N 18341-001
Conductive Silicon Gasket P/N 10882
Loctite 4541 Adhesive P/N 10773
43
Back Panel
If not previously installed:
Tee Fitting P/N 18186-001
NOTE
If you have not performed the calibrations shown in Chapter 6 you will need to measure the
turbine speed as shown on page 6-22 before replacing the back panel. This is required in
order to complete the Final Checkout tests in Chapter 9.
1)
Before replacing the back panel, verify that any DIP switch settings have been restored
to their correct positions as shown below.
2)
Turn the ventilator face down.
3)
Check the serial number on the back of the ventilator to ensure it matches the number
found on the inside of the ventilator near the fan assembly or at the top of the inlet foam
filter housing. If the serial numbers do not match, locate the correct back panel for the
ventilator before continuing.
43
Contact CareFusion for back panel replacement information and proper labeling information.
44
O
Sampling Tube, P/N 10544, ~10.0" long, 0.125" O.D. X 0.079 I.D. clear polycarbonate tubing.
2
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with torque
meter
Grounded anti-static wrist strap
O
Sampling Tube, P/N 10544
2
44