CareFusion LTV 1200 Service Manual page 294

Table of Contents

Advertisement

Appendix A - V
Modes and Breath Types
Breath Types
Modes
Variable Controls
Control
Backup Pressure Trigger
Breath Rate
Date Format
Display Select
Inspiratory/Expiratory
Hold
Inspiratory Time
Leak Compensation
Language
®
O
% (LTV
1200 Only)
2
®
O
Flush (LTV
1200
2
Only)
p/n 18603-001, Rev. J
ENTILATOR
Volume Control, Pressure Control, Pressure Support, Spontaneous
Control, Assist/Control, SIMV, CPAP, NPPV, Apnea Backup
Range
-3 cmH
O
2
"--", 1 to 80 bpm
mm/dd/yyyy, dd/mm/yyyy,
yyyy/mm/dd
Toggles between manual or
automatic display scrolling
and changes monitor
displayed.
One push toggles monitor
window display between
normal display, INSP HOLD
and EXP HOLD.
While INSP HOLD is
displayed, a push and hold
initiates an Inspiratory Hold.
While EXP HOLD is
displayed, a push and hold
initiates an Expiratory Hold.
0.3 to 9.9 seconds
On, Off
English, Dansk, Deutsch,
Español, Francais, Italiano,
Norsk, Portugues, Svenska,
Русско
21% to 100%
O
:
95%
2
Time:
1, 2, or 3 minutes
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
S
PECIFICATIONS
Tolerance
 2 cmH
 1 bpm or 10% of breath period,
whichever is less
n/a
n/a
6 seconds maximum
6 seconds maximum
 0.05 seconds
n/a
n/a
O
% mean: 21% to 50%:
2
 5
 0.1 sec
O
2
 3
 5
51% to 95%:
keep steady-state only
Page A-1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100