Power Checkout Worksheet - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Power Checkout Worksheet

®
LTV
SERIAL NUMBER:
TEST
DESCRIPTION
Battery Level Test
Replace battery with the 10A DC
voltage source set between 14.8V and
15.0V and turn unit on. Clear the power
lost alarm.
Slowly reduce voltage until battery LED
changes to amber.
Slowly reduce voltage until battery LED
changes to red.
Slowly reduce voltage until Battery
Level LED extinguishes.
Remove DC voltage source and
reconnect the internal battery.
Battery Duration/Battery Charge Test
Run unit from internal battery until it
shuts down. Charge from external
power while operating for a minimum of
8 hours.
Remove external power and operate
unit from internal battery. Use settings:
Mode: Volume, Assist/Ctrl
Breath Rate: 15
Tidal Volume: 800
Insp. Time: 1.5 sec
PEEP: 5
Disconnect the internal battery, re-
connect external AC adapter. Clear
Reset alarm.
Reconnect the internal battery.
Page 9-16
____________
STEP
#
1)A)
1)B)
1)C)
1)D)
1)E)
2)A)
2)B)
2)C)
2)D)
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
CONDUCTED BY:
DATE:
REQUIREMENT
Normal operation
Battery Level LED is green
Only POWER LOST alarm activates
Normal operation
DC voltage source is 11.9V  0.238 (2%)
BAT LOW and Audible alarm
Normal operation
DC voltage source is 11.5V  0.230 (2%)
BAT EMPTY and Audible alarm
DC voltage source is 10.0V  0.200 (2%)
Vent Inop LED illuminated.
Audible alarm sounds.
Vent Inop LED lights and audible alarm
sounds
Battery Level LED is green upon
removing external power.
Normal operation from internal battery for
minimum of 40 minutes.
Charge Status LED is red
Charge Status LED flashes amber for a
few minutes, and then goes solid amber.
____________
____________
PASS
(P) /
FAIL
(F)
p/n 18603-001, Rev. J