26) Reconnect the three (3) flexible tubes to the Flow Xducer and Exh Valve fittings on the
side panel of the ventilator.
27) Reinstall the flow valve assembly (see instructions on page 8-66).
28) Reconnect the flexible tubes to the solenoid manifold following the internal flexible tube
routing table, instructions and diagrams beginning on page 8-33. Inspect all the
flexible tubes for tears at the connecting ends and replace worn or damaged
tubes if necessary.
NOTE
Tubing Configurations - When disconnecting tubes, note their positions and review the
internal flexible tube routing tables, instructions and diagrams beginning on page 8-33.
If tubing needs to be replaced, consult the tables and instructions in Internal Flexible Tubing
beginning on page 8-33 for tubing diameters, lengths and routing diagrams.
29) Reconnect all flexible tubing to the analog board. See instructions on page 8-57.
30) Reconnect the 3-wire and 5-wire connectors from the turbine to the motor board. The
connectors are keyed and will snap into place when properly connected.
31) Reconnect the 3-wire flow valve connector to the power board. The connector is keyed
and will snap into place when properly connected.
32) Reconnect the Pisco connector to the oxygen pressure transducer on the analog board
®
(LTV
1200 only).
33) Reconnect the 4-wire flow valve connector to the motor board and the 2-wire flow valve
temperature connector to the power board. The connectors are keyed and will snap into
place when properly connected.
34) Install an O
Sampling Tube
2
Component Removal and Replacement for detailed instructions.
35) Verify that no tubing is kinked or twisted, and that it will not be pinched when
compressed by the back panel. Refer to the routing instructions, tables and diagrams
beginning on page 8-33 to check the tubing routing. Always check ALL tube routing
for kinks or compression that might result in restricted flows prior to closing the
ventilator.
36) Reconnect the internal battery and replace the back panel (see page 8-29).
64
O
Sampling Tube, P/N 10544, ~10.0" long, 0.125" O.D. X 0.079 I.D. clear polycarbonate tubing.
2
Page 8-116
64
. See Back Panel, Reinstallation in Chapter 8 –
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J