Solenoid Manifold, Analog PCBA and External Port Tubing Installation:
NOTE
Some previously attached manifold tubing is not shown for clarity of pertinent instructions.
1)
Attach Solenoid Manifold port tubing, as follows;
port # 4 onto ventilator side panel Low Pressure port
port # 8 over tubing from port # 4 and onto ventilator side panel High Pressure
port
2)
Insert the elbow fitting attached to the Tee and Elbow subassembly on Solenoid
Manifold Port # 1 into the tubing on the High Pressure port on the Valve Differential
Transducer.
the Tee and Elbow subassembly is routed under the Flow Differential
Transducer and PEEP Pilot Pressure Transducer ports and over the Tube and
Tee subassembly from Solenoid Manifold side port # 9