Recommended Maintenance Schedule - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Chapter 5 - P
The following preventative maintenance is required for proper operation of the LTV
1100 ventilator.

Recommended Maintenance Schedule

®
The LTV
1200, 1150, and 1100 ventilator is designed to operate for extended periods of time with
minimum maintenance. The following maintenance is recommended:
8
Hours of Service
Prior to initial use
While in storage, every
two months
Daily
If in use, a minimum of
once a month
Every 10,000 hours or
10
two years
, whichever
comes first
Every 30,000 hours or
six years, whichever
10
comes first
This is the recommended schedule for typical clinical or home settings. Some environmental
conditions may require you to perform the maintenance procedures more frequently.
8
For the number of hours the ventilator has been in service, see Extended Features in the LTV Operator's Manual.
9
If the battery is deeply discharged, it may take several charge and discharge cycles before it is at full capacity.
10
10,000 hour, two year and/or 30,000 hour, six year extended maintenance and ventilator repair must be performed by a
service technician who has been trained and certified by CareFusion.
11
Replacement at 10,000 hours or 2 years is based on normal use of up to 200 charge cycles. The battery may need to be
replaced more frequently if it is being charged more often. The battery should also be replaced any time it fails to reach a
full charge, or if the ventilator runs for less than 40 minutes on a fully charged battery.
12
®
The LTV
1200, 1150, and 1100 Internal Battery is available in battery replacement kit, P/N 18634-001.
13
The Turbine Manifold Assembly (part no. 11490) must be replaced after 30,000 hours of operation. A Turbine Manifold
Assembly six years or older with fewer than 30,000 hours may continue operation, as long as it passes the Performance
Checkout Test (found in the LTV
hours). All other maintenance within the 30,000 hours, or six years of operation, is required at the specified service interval.
p/n 18603-001, Rev. J
REVENTATIVE
Maintenance Required
Charge the internal battery by plugging the ventilator into an AC power source for
24 hours
Setup the ventilator/accessories
Check the ventilator for proper operation per Chapter 2 - Ventilator Checkout
Tests
Recharge the internal battery by plugging the ventilator into an AC power source
9
for 24 hours
Check the Inlet filter, clean if necessary
Check the Fan filter, clean if necessary
Check the ventilator per Chapter 2 - Ventilator Checkout Tests. While the
ventilator is off-patient, perform the power (external) disconnect test*
Verify Vte or VE monitor *
Verify airway pressure or PIP monitor *
Verify delivered O
concentration (LTV
2
analyzer continuously. * (See Ventilator Specifications for accuracy tolerances)
*Use existing patient settings or examples shown in Installation and Checkout
Replace the Internal Battery
Calibrate the Transducers
Replace the Power Board
Clean or replace the Interior Air Inlet filter
Clean or replace the O
Replace the Alarm Sounder Assembly
Replace the Turbine Manifold assembly
Replace the Solenoid Manifold
Replace the Flow Valve
Replace the Rotary Switch assembly
Replace the O
Blender (LTV
2
Replace the Solenoid Mount assembly
Replace the PEEP Accumulator
Replace the Fan assembly
Replace all Silicone Tubing
Check the Thermo Pads for compression and replace if necessary
®
service manual) at six years and every two years thereafter (until reaching 30,000
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
M
AINTENANCE
®
1200 only) if not using an oxygen
11
with CareFusion battery P/N 18608-001
®
Inlet filter (LTV
1200 only)
2
13
®
1200 only)
®
1200, 1150, and
12
Page 5-1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents