Symptoms
CONTINUED ...
Repeated SBT < f
alarms.
Repeated SBT > f/Vt
alarms.
Repeated SBT < f/Vt
alarms.
p/n 18603-001, Rev. J
Possible Causes
Internal problem with the
ventilator.
Total Breath Rate (f)
exceeds the set SBT > f/Vt
alarm value.
Patient Circuit leak.
Internal problem with the
ventilator.
Total Breath Rate (f) is less
than the set SBT < f/Vt
alarm value.
Patient Circuit leak.
Internal problem with the
ventilator.
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
What to Do
If problem reoccurs, immediately contact a
certified CareFusion service technician.
Check SBT > f/Vt alarm value. (See the
®
LTV
1200, 1150, or 1100 Ventilator
Operator's Manual for instructions).
Run a Leak test and reseat or replace the
leaking parts or connections. (See the
®
LTV
1200, 1150, or 1100 Ventilator
Operator's Manual for instructions).
If problem reoccurs, immediately contact a
certified CareFusion service technician.
Check SBT < f/Vt alarm value. (See the
®
LTV
1200, 1150, or 1100 Ventilator
Operator's Manual for instructions).
Run a Leak test and reseat or replace the
leaking parts or connections. (See Leak
Test in Chapter 2 for instructions).
If problem reoccurs, immediately contact a
certified CareFusion service technician.
Page 7-39