Final Tubing and Cable Installation:
1)
Route the tube from Solenoid Manifold port # 3 under the tube and tee attached to
Solenoid Manifold port # 7, over previously installed tubing and onto the Airway
Pressure Transducer port, as shown.
2)
Route the tube from Solenoid Manifold port # 2 under the tube and tee attached to
Solenoid Manifold port # 7, over the previously attached tubing, through the notch
in the Analog PCBA and onto the Flow Differential Transducer port, as shown.
do not loop past the edge of the Power PCBA into the area for the Internal
Battery
3)
Route the 3-wire, Flow Valve cable from the Flow Valve over previously installed
tubing and the Accumulator. Attach the connector to the 3-wire connector (JP2) on
the Power PCBA, as shown. The connector is keyed to fit only one way and will snap
into place when properly aligned.
Page 8-50
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J