13) Remove the power board mounting screws (2 attaching the power board and/or the
Pigtail cable assembly to the side panel from the outside (retain) and five (5) located on
the face of the board). See illustration below.
14) Pull the pigtail connector interface bracket straight up and off the power board
connectors and the molded strain relief portion of the pigtail cable assembly.
15) Disconnect the 4-wire pigtail cable assembly connector from JP6 on the power board
and slide the pigtail cable assembly out of the keyed-round hole of the upper weldment
and set aside.
16) Disconnect the two 2-wire connectors from the purge solenoids and remove the
mounting nut from the solenoid manifold.
17) Insert a power PCB separator tool (P/N 11599) into the gap between the edge of the
power board and the turbine manifold, at the approximate mid-point of the 64-pin
connector (see Detail illustration on next page). The tool may need to be moved left or
right along the gap to insert.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-107