Internal Battery Pack - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Internal Battery Pack

Parts required for replacement:
 LTV
®
1200, 1150, and 1100 Internal
Replacement Battery P/N 18634-001.
CAUTION
Internal Battery Use - The internal battery is intended for use in LTV
short periods while switching between external power supply connections, emergency
situations or short duration transports. The length of time the ventilator will operate on
battery power is a function of many factors such as settings, charge level and condition or
age of the battery; therefore, the use of the internal battery as a standard operating practice
is not recommended.
To remove and replace the battery pack:
1) Position the LTV
clean, dry work surface and use a 3/32" hex bit to remove both button socket head
cap screws (2, p/n 18296-001) from the Battery Cover.
2) Open the cover and take note of how the battery connector wires are positioned
between the side of the case and the battery.
p/n 18603-001, Rev. J
®
1200, 1150, or 1100 ventilator as shown (front side down) on a
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Torque driver (120 in-oz/0.84 Nm)
3/32" Hex bit
®
ventilators during
Page 8-81

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents