Alarm Sounder Assembly (current version)

Parts required for replacement:
Sounder Assembly P/N 21150-001
(contained in Sounder Kit P/N 21677-001)
Replace if damaged:
1/4" Pan-head Screw (1) P/N 10435
To replace the Sounder Assembly:
1) Remove the ventilator back panel and disconnect the internal battery cable from the
power board (see instructions on page 8-28). Remove the left and right soft side panels.
2) Disconnect the 2-wire sounder connector from the power board.
3) Remove the 1/4" pan-head screw from the sounder bracket (previous version) or
sounder assembly (current version) as shown.
4) Remove the sounder and sounder bracket (previous version) or sounder assembly
(current version).
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with
torque meter
Grounded anti-static wrist
strap
Page 8-57