Leak Test

The leak test is used to test the patient circuit for leaks.
To run the leak test:
1) At the end of the control test, LEAK shows in the LED display window.
2) Attach a patient circuit with all accessories (such as water traps, heated circuits and
humidifiers to the patient circuit).
3) Connect the patient circuit to the ventilator.
4) With a clean, gloved hand or 4"X4" gauze pad, occlude the proximal end of the patient
circuit.
5) Press Select while
NOTE
The leak test cannot be run until the ventilator has been running for 60 seconds. If you
attempt to run the leak test before the warm-up period has completed, a WARMUP xx
message will be displayed. When the warm-up period is complete, the leak test menu is
redisplayed.
®
Although the LTV
and 1100 Ventilators.
Page 2-8
LEAK
is displayed.
1200 Front Panel is shown, the test is applicable to the LTV
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
1150
p/n 18603-001, Rev. J