The AVEA
®
ventilator's quick tips card is not intended as a replacement for the
operator's manual.
You must become completely familiar with the operator's manual before using
the AVEA ventilator.
Touch-Turn-Touch
1.
Touch the control to select. The control changes
color indicating that it is active.
2.
Turn the data dial to reach the selected value.
3.
Touch the control again or press
confirm the change.
Setting alarm limits
1.
Press the

ALARM LIMITS

close the window.
2.
Use Touch-Turn-Touch or
Touch-Turn-Accept
method to modify alarms.
Proximal flow sensors
The AVEA ventilator accepts either hot wire or variable orifice proximal low
sensors. These are in addition to the internal inspiratory flow sensor and heated
expiratory flow sensor.
To attach, pull back the locking collar. Push firmly onto the receptacle. Push the
collar forward to lock the sensor in place.
To disconnect, retract the collar and pull the connector straight
away from the ventilator.
and Touch-Turn-Accept
ACCEPT
button to open or
2
The variable orifice sensor attaches to the receptacle
circled in dark blue.
The hot wire flow sensor attaches to the receptacle
circled in light blue directly below the variable orifice
flow sensor connection.
method
1
2
3
to
1