AVEA ventilator systems

Analog Input Configuration

Under the heading "Set Analog Input Scale" there are two buttons representing two possible voltage ranges.
If the full-scale output of the device you are interfacing with is less than 1 volt, select the 0-1 volt scale button.
If it is 5V or less, select the 0-5 volt range. Select the appropriate analog scale and press the ACCEPT key to enter the
configuration.
Analog Input is configured on the same connector as the ILV. The pin configuration for cables to use this feature is
shown in Figure 2–36 below. Pin configuration of the connector for attachment to your other device must be supplied by
the manufacturer of that device
WARNING
All applications using this connector require specially made cables. DO NOT connect a standard DB25 cable
to this receptacle. This could result in damage to the ventilator. Contact Technical Support at the numbers
given in Appendix A.
Analog Input Channel 0
Figure 2–36: Analog Input connections
L2786 Rev. M
Chapter 2: Unpacking and Setup
Analog Input Channel 1
33