PD1500A Double-Pulse Test System Calibration
Gate Voltage Source Calibration – V
Hardware Setup:
Figure 7
Setup for Gate Voltage Calibration
Procedure:
After setting up the hardware as described above, click Start to perform the
calibration and continue to the Gate Current Calibration.
66
Refer to Figure 7 below.
– Remove the I/NO Jumper on the Low-Side Gate Drive Module.
– Install the V/NC Jumper on the Low-Side Gate Drive Module.
– Connect the BNC to Banana Adapter to the BNC to SMA calibration cable.
– Connect the banana plug cable to Channel 1 Force terminals of the
B2902A SMU (2-wire measurement)
– Connect the SMA end of the cable to the SMA connector on the Low-Side
Gate Drive Module.
– The Safety Enclosure hood may be left open.
1251-2277
Banana - BNC
Adapter
Low-Side
Gate Drive
Module
A strip of 10 header jumpers (Keysight part number 1258-0225)
were supplied with your PD1500A for use during calibration.
System Calibration Procedure (Autocal)

Output

GS
8121-2816
BNC - SMA
Cable
I/NO (Normally Open) Jumper.
Remove the jumper for Gate
Voltage Source Calibration.
V/NC (Normally Closed)
Jumper
Keysight Double-Pulse Test System Installation Guide