DC Attenuation/Offset Calibration Using N7022A Main Cable

1
Turn on the oscilloscope and connect the N7020A probe with the N7022A main
cable attached to one of the oscilloscope's input channels.
2
Allow the oscilloscope and probe to warm up for 20 minutes.
3
If the oscilloscope needs calibration, perform a user calibration before the probe
calibration. On the oscilloscope, click Utilities > Calibration.
4
Connect the output of the N7022A main cable to the oscilloscope's rear-panel
AUX OUT BNC connector as shown in the following picture. You'll need a BNC (m)
to SMA (f) adapter.
Figure 11
5
On the oscilloscope, click Setup > Probes.
6
In the Probe Calibration dialog box, select the tab representing the channel that
has the probe attached.
7
In the dialog box, select the type of calibration. Click Start and follow the
instructions shown on the oscilloscope.