Compatible SCPI Commands
Supported SCPI Commands
State Saved
CALCulate<1|2>:LIMit<1...6>:ACPower:ALTernate<1|2>:ABSolute
Specifies the absolute limit value for the lower/upper alternate adjacent channel power measurement
(Adjacent Channel Power) in the selected measurement window.
The numeric suffix after ALTernate<1|2> denotes the first or the second alternate channel. The numeric
suffixes <1 to 6> are irrelevant for this command.
Remote Command
Example
Notes
Preset
State Saved
CALCulate<1|2>:LIMit<1...6>:ACPower:ALTernate<1|2>:ABSolute:STATe
Specifies the limit check for the first/second alternate adjacent channel in the selected measurement
window for ACP measurement.
Before the command, the limit check must be globally switched on for the channel/adjacent channel
power with the command CALCulate:LIMit:ACPower:STATe ON.
The numeric suffix after ALTernate<1|2> denotes the first or the second alternate channel. The numeric
suffixes <1 to 6> are irrelevant for this command.
The result can be queried with CALCulate:LIMit:ACPower:ALTernate<1|2>:RESult?. A complete
measurement must be performed between switching on the limit check and the result query, since
otherwise no valid results are available.
Remote Command
Example
Notes
Preset
State Saved
76
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
Saved in instrument state.
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ALTernate[1]|2
:ABSolute <real>,<real>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ALTernate[1]|2
:ABSolute?
CALC:LIM:ACP:ALT:ABS 30DBM, 30DBM
CALC:LIM:ACP:ALT:ABS?
FSU, FSP, FSE
–200dBm, –200dBm
Saved in instrument state.
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ALTernate[1]|2
:ABSolute:STATe ON|OFF|1|0
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ALTernate[1]|2
:ABSolute:STATe?
CALC:LIM:ACP:ALT:ABS:STAT ON
CALC:LIM:ACP:ALT:ABS:STAT?
FSU, FSP, FSE
OFF
Saved in instrument state.