Installation Note
Agilent MXA Signal Analyzer
Front Frame Assembly Replacement
Part Number N9020-90006
Printed in USA October 2006

Advertisement

loading

Summary of Contents for Agilent Technologies MXA N9020A

  • Page 1 Installation Note Agilent MXA Signal Analyzer Front Frame Assembly Replacement Part Number N9020-90006 Printed in USA October 2006...
  • Page 2 The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Front Frame Assembly MXA Installation Kit Front Frame Assembly MXA Installation Kit Products Affected: MXA N9020A Serial Numbers: To Be Performed By: (X) Agilent Service Center (X) Personnel Qualified by Agilent (X) Customer Estimated Installation Time: 2.0 Hours Functional Check:...
  • Page 4 Introduction This replacement kit provides everything needed to replace the Front Frame on an Agilent MXA N9020A Signal Analyzer. If the existing Frequency Range Label can not be re-used then a replacement label must be ordered separately. Frequency Range Label Part Numbers As described in the Service Guide, the Frequency Range Labels are not shipped with Front Frame Assembly kits.
  • Page 5: Installation Procedure

    Front Frame Assembly MXA Installation Kit Installation Procedure Standard Instrument (Benchtop Configuration) NOTE Make sure any connectors on the front panel are removed. 1. Disconnect the instrument from ac power. 2. Refer to Figure 1. Using the T-20 driver, remove the 4 screws (1) (two on each side) that attach the strap handles (2) on each side of the instrument.
  • Page 6 Front Frame Assembly MXA Installation Kit Option PRC Instrument (Portable Configuration) NOTE Make sure any connectors on the front panel are removed. 1. Disconnect the instrument from ac power. 2. Refer to Figure 2. Using the T-20 driver, remove the four screws (1), two on each side, to detach the Bail Handle Assembly (2) from the front frame.
  • Page 7 Front Frame Assembly MXA Installation Kit Figure 3 Option PRC Instrument Outer Case Removal Installation Note N9020-90006...
  • Page 8 Front Frame Assembly MXA Installation Kit Front Frame Assembly Removal 1. Refer to Figure 4. Using the T-10 driver, remove the eight screws (1), four on each side, to detach the front frame from the chassis. 2. Refer to Figure 5.
  • Page 9 Front Frame Assembly MXA Installation Kit 3. Refer to Figure 6. Remove the RPG knob (17) by carefully pulling it off. Figure 6 Front Frame Exploded View Installation Note N9020-90006...
  • Page 10 Front Frame Assembly MXA Installation Kit 4. Refer to Figure 7. Ribbon cable W1 (26) must be separated from the back of the display by pulling up to separate the velcro that is used to adhere the cable to the bracket.
  • Page 11 Front Frame Assembly MXA Installation Kit 12.It is going to be difficult to reuse the Frequency Range label from the existing front frame. The Service Guide referencing this kit recommended that the appropriate Frequency Range label also be purchased. 13.Set aside the existing front frame as a reference for later steps. Refer to Figure 6 when assembling the remaining items into the new...
  • Page 12 Front Frame Assembly MXA Installation Kit Final Installation for Standard Instruments 1. Locate the two vinyl Trim - sides from the kit (5041-9172). Peel off the self adhesive backing and apply to the sides of the instrument front frame. Use the old front frame as a reference.
  • Page 13: Functional Check

    If you require further assistance, please contact the Agilent support team. Online assistance: http://www.agilent.com/find/assist If you do not have access to the Internet, contact your local Agilent Technologies Sales and Service Office, or if in the United States, call 1-800-829-4444 (8 am - 8 pm ET, Monday - Friday).
  • Page 14 Front Frame Assembly MXA Installation Kit Installation Note N9020-90006...