Post-Repair Procedures
What You Will Find in This Chapter
Programming Model and Serial Numbers
When the A7 Midplane board assembly is replaced the instrument model and serial
numbers must be reprogrammed into the instrument. To program the instrument
with this information a software utility must be downloaded from the following
internet site:
http://sa.tm.agilent.com/N9039A/midplane/
Information on the instrument software requirements as well as instructions for
using this utility can be found on this web site as well.
This utility will only allow the writing of the instrument model and serial numbers
into an instrument that does not already have valid numbers in it. Since the
instrument model and serial numbers are stored on the A7 Midplane board
assembly, this should only happen when this assembly has been replaced with a
new one.
Chapter 8
231