Installation Note
Agilent MXA Signal Analyzer
Option MS1, Upgrade to 2 GB SDRAM
Part Number N9020-90047
Printed in USA September 2007

Advertisement

loading

Summary of Contents for Agilent Technologies MXA Series

  • Page 1 Installation Note Agilent MXA Signal Analyzer Option MS1, Upgrade to 2 GB SDRAM Part Number N9020-90047 Printed in USA September 2007...
  • Page 2 The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Upgrade to 2 GB SDRAM Kit Upgrade to 2 GB SDRAM Kit Products Affected: MXA N9020A Serial Numbers: To Be Performed By: (X) Agilent Service Center (X) Personnel Qualified by Agilent (X) Customer Estimated Installation Time: 1.0 Hours Estimated Adjustment and Verification Time: 1.0 Hours Additional Tasks: Installation Note...
  • Page 4 Upgrade to 2 GB SDRAM Kit Introduction This installation note contains the instructions needed to upgrade the analyzer to 2 GB SDRAM. 2 GB SDRAM Installation Procedure This installation note is part of a retrofit kit that provides all of the parts and instructions needed to upgrade the analyzer to 2 GB of total SDRAM.
  • Page 5: Installation Procedure

    Upgrade to 2 GB SDRAM Kit Installation Procedure Electrostatic discharge (ESD) can damage or destroy electronic CAUTION components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
  • Page 6 Upgrade to 2 GB SDRAM Kit 2. Refer to Figure 2. Remove and discard the 15 machine screws (1) from the CPU assembly. The screws need to be discarded and replaced with the 15 screws provided in this kit. NOTE Do not remove the 4 Hard Drive screws.
  • Page 7 Upgrade to 2 GB SDRAM Kit 3. Carefully lift up the CPU top shield and unlock the Hard Drive ribbon cable from the CPU Board by pressing down and out on the two locking tabs located on the sides of the connector as shown in Figure Figure 3...
  • Page 8 Upgrade to 2 GB SDRAM Kit 4. Set aside Top Shield/Hard Drive/Ribbon Cable assembly. Figure 4 Top Shield/Hard Drive Assembly 5. Locate the 1 GB SDRAM board assembly from the kit. See Figure Figure 5 1 GB SDRAM Board 6. Before installing the SDRAM board, use a high pressure non-CO2 air duster to clean off the CPU board and the vacant connector in case there is dried loctite from the 15 screws that were removed earlier.
  • Page 9 Upgrade to 2 GB SDRAM Kit Figure 6 Insert the 1 GB SDRAM Board 8. Connect the Hard Drive ribbon cable from the Top Shield/Hard Drive/Ribbon Cable assembly from step 4 to the CPU. When the ribbon cable is fully seated the two locking tabs will grip the sides of the ribbon cable connector.
  • Page 10: Functional Check

    Displayed Average Noise Level Residual Responses End of installation. For assistance, contact your nearest Agilent Technologies Sales and Service Office. To find your local Agilent office access the following URL, or if in the United States, call the following telephone number: http://www.agilent.com/find/assist...