Compatible SCPI Commands
Supported SCPI Commands
The query returns the current value set.
This mode does not support any unit for this command.
Remote Command
Example
Remote Command Notes
State Saved
CALCulate<1|2>:LIMit<1...6>:LOWer:STATe ON | OFF
The command activates or deactivates the specified lower limit line. Limit check is activated separately,
using CALC:LIM:STAT ON.
The query returns the current state of this setting.
Remote Command
Example
Remote Command Notes
Preset
State Saved
CALCulate<1|2>:LIMit<1...6>:LOWer:OFFSet <numeric_value>
The command defines an offset for the Y axis of the specified relative lower limit line.
This mode does not support any unit for this command.
The query returns the current value of this setting.
Remote Command
Example
Remote Command Notes
Preset
80
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer[:DATA]
<real>,<real>...
CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer[:DATA]?
CALC:LIM:LOW:DATA 1,20,100,300,1
CALC:LIM:LOW:DATA?
FSE, FSP, FSU
Saved in instrument state.
:CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:STATe ON|OFF|1|0
:CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:STATe?
CALC:LIM:LOW:STAT OFF
CALC:LIM:LOW:STAT?
FSE, FSP, FSU
OFF
Saved in instrument state.
:CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:OFFSet <real>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:OFFSet?
CALC:LIM:LOW:OFFS 1
CALC:LIM:LOW:OFFS?
FSE, FSP, FSU
0