[Sense<1|2>:]Power:achannel:acpairs 0 | 1 | 2 | 3; [Sense<1|2>:]Power:achannel:bandwidth|Bwidth[:Channel] 100 Hz To 1000 Mhz - Agilent Technologies X Series User Manual

Signal analyzer, emi receiver
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

adjacent-channel power measurements.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Min
Max
[SENSe<1|2>:]POWer:ACHannel:ACPairs 0 | 1 | 2 | 3
Specifies the number of adjacent channels (upper and lower channel in pairs).The number 0 represents a
pure channel power measurement.
Remote Command
Example
Remote Command Notes
Preset
Range
[SENSe<1|2>:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel] 100 Hz to 1000 MHz
Specifies the radio communication system's channel bandwidth.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Min
Max
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
[:SENSe]:POWer:ACHannel:TXCHannel:COUNt <int>
[:SENSe]:POWer:ACHannel:TXCHannel:COUNt?
POW:ACH:TXCH:COUN 1
POW:ACH:TXCH:COUN?
FSP, FSU
4
Saved in instrument state.
1
4
[:SENSe]:POWer:ACHannel:ACPairs
[:SENSe]:POWer:ACHannel:ACPairs?
POW:ACH:ACP 3
FSE, FSP, FSU
0
0|1|2|3
[:SENSe]:POWer:ACHannel:BANDwidth|BWIDth[:CHANnel]
<freq>
[:SENSe]:POWer:ACHannel:BANDwidth|BWIDth[:CHANnel]?
POW:ACH:BWID 30kHz
FSE, FSP, FSU
14KHz
Saved in instrument state.
100Hz
1000MHz
Compatible SCPI Commands