Compatible SCPI Commands
Agilent N9062A/W9062A Supported IEEE SCPI Commands
Table 3-2
SCPI Commands
*RST
*SAV <register>
*SRE <integer>
*SRE?
*STB?
*TRG
*TST?
*WAI
56
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
IEEE 488.2 SCPI Commands
Description
This command presets the instrument to a factory defined condition that is appropriate for
remote programming operation. *RST is equivalent to performing the commands:
1.:SYSTem:PRESet, with preset type set to MODE.
2.*CLS which clears the STATus bits and error queue
3.:INITiate:CONTinous OFF which selects single sweep/measurement
*RST does not change the mode and only resets the parameters for the current mode.
This command saves the instrument state to the specified instrument memory register.
This command sets the value of the service request enable register.
The query returns the value of the register.
Returns the value of the status byte register without erasing its contents.
This command triggers the instrument. Use the command :TRIGger[:SEQuence]:SOURce
to select the trigger source.
This query performs a self test and returns a number indicating the success of the testing.
A zero is returned if the test is successful.
This command causes the instrument to wait until all pending commands/processes are
completed before executing any additional commands.
There is no query form for the command.