Compatible SCPI Commands
Supported SCPI Commands
Preset
State Saved

TRIGger Subsystem

This subsystem includes commands that control and synchronize the start of a sweep.
An external trigger signal can be supplied via the connector on the analyzer's rear panel.
TRIGger<1|2>[:SEQuence]:SOURce
Selects the trigger source.
This mode supports only parameter options: IMMediate, EXTernal, VIDeo, RFPower.
It does not support parameter options: IFPower, TV, AF, FM, AM and PM.
Remote Command
Example
Remote Command Notes
Remote Command Notes
Preset
Range
TRIGger<1|2>[:SEQuence]:LEVel:IFPower –30 to –10DBM
This command sets the level of the IF power trigger source.
This command is accepted, but takes no action and reports no error.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Min
Max
TRIGger<1|2>[:SEQuence]:LEVel:RFPower
Sets the RF power trigger level.
162
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
OFF
Saved in instrument state.
:TRIGger[1][:SEQuence]:SOURce
IMMediate|EXTernal|VIDeo|IFPower|RFPower|TV|AF|FM|AM|PM
:TRIGger[1][:SEQuence]:SOURce?
TRIG:SOUR EXT
FSE, FSP, FSU
IMMediate
IMMediate|EXTernal|VIDeo|IFPower|RFPower|TV|AF|FM|AM|PM
:TRIGger[1][:SEQuence]:LEVel:IFPower <ampl>
:TRIGger[1][:SEQuence]:LEVel:IFPower?
TRIG:LEV:IFP –20DBM
FSP, FSU
–20dBm
Saved in instrument state.
30dBm
–70dBm