Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:MARKer<1...4>:FUNCtion:NOISe:RESult?
Queries the noise measurement result.
To obtain a valid result, a complete sweep must be performed after switching on the function, but before
querying the measured value. This is possible only in Single sweep mode.
Remote Command
Example
Remote Command Notes
CALCulate<1|2>:MARKer<1...4>:FUNCtion:REFerence?
Sets the reference level to the power measured by the specified marker.
If marker 2, 3 or 4 is selected, and used as delta marker, it is switched to marker mode.
This command is an event, and therefore has neither *RST value nor query.
The value of the numeric suffix <1 to 4> is ignored.
Remote Command
Example
Remote Command Notes
CALCulate:MARKer:FUNCtion:POWer Subsystem
The CALCulate:MARKer:FUNCtion:POWER subsystem includes commands for control of power
measurement.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:SELect ACPower | CPOWer |
MCACpower | OBANdwidth | OBWidth | CN | CN0
Selects and activates the specified power measurement type.
This mode supports: ACPower, MCACpower, CPOWer, OBANdwidth and OBWidth.
Remote Command
Example
Remote Command Notes
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:RESult? ACPower | CPOWer |
MCACpower | OBANdwidth | OBWidth | CN | CN0
Queries the result for the specified power measurement type.
96
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
:CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:NOISe:RESult?
CALC:MARK:FUNC:NOIS:RES?
FSE, FSP, FSU
:CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:REFerence
CALC:MARK:FUNC:REF
FSE, FSP, FSU
CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:POWer:SELect
ACPower|CPOWer|MCACpower|OBANdwidth|OBWidth|CN|CN0
CALC:MARK:FUNC:POW:SEL OBW
FSE, FSP, FSU