[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution] <numeric_value>
Specifies the instrument's resolution bandwidth.
In the SCPILC mode, the maximum resolution bandwidth is 8 MHz (not 10 MHz as in FSx) and the
minimum resolution bandwidth is 1 Hz (not 10 Hz as in FSx).
Remote Command
Example
Remote Command Notes
Remote Command Notes
Remote Command Notes
Preset
State Saved
Min
Max
Backwards Compatibility SCPI
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:AUTO ON | OFF
Automatically couples the instrument's resolution bandwidth to the span or it cancels the coupling.
Remote Command
Example
Remote Command Notes
Preset
Backwards Compatibility SCPI
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:RATio 0.0001 to 1
Specifies the resolution bandwidth (Hz) / span (Hz) ratio.
Remote Command
Example
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
BAND 1 KHZ
BAND?
The setting and querying of values depends on the current bandwidth type.
FSE, FSP, FSU
3 MHz
Saved in instrument state.
1 Hz
8 MHz
:SENSe1:BANDwidth|BWIDth[:RESolution]
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
BWID:AUTO ON
BWID:AUTO?
FSE, FSP, FSU
ON
:SENSe1:BANDwidth|BWIDth[:RESolution]:AUTO
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio <real>
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio?
BAND:RAT 0.1
Compatible SCPI Commands