Compatible SCPI Commands
Supported SCPI Commands
Example
Remote Command Notes
Preset
CALCulate<1|2>:LIMit<1...6>:UPPer:MARGin <numeric_value>
The command defines a margin for the specified upper limit line. At the margin level, out-of-limit values
are signaled (if the limit check is active), but are not treated as violations of the limit value.
This mode does not support any unit for this command.
The query returns the current value of this setting.
Remote Command
Example
Remote Command Notes
Preset
CALCulate<1|2>:LIMit<1...6>:UPPer:MODE RELative | ABSolute
The command switches between relative or absolute scaling for the Y-axis of the specified upper limit
line.
Selecting RELative causes the unit to be switched to dB.
The query returns the current state of this setting.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Range
CALCulate<1|2>:LIMit<1...6>:UPPer:SHIFt <numeric_value>
The command shifts the specified limit line in the Y-direction by the specified amount.
84
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
CALC:LIM:UPP:OFFS 1
CALC:LIM:UPP:OFFS?
FSE, FSP, FSU
0
:CALCulate[1]:LIMit[1]|2|3|4|5|6:UPPer:MARGin <real>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:Upper:MARGin?
CALC:LIM:UPP:MARG 10
CALC:LIM:UPP:MARG?
FSE, FSP, FSU
0
:CALCulate[1]:LIMit[1]|2|3|4|5|6:UPPer:MODE
RELative|ABSolute
:CALCulate[1]:LIMit[1]|2|3|4|5|6:UPPer:MODE?
CALC:LIM:UPP:MODE ABS
CALC:LIM:UPP:MODE?
FSE, FSP, FSU
ABS
Saved in instrument state.
RELative | ABSolute