TRIGger Subsystem

The TRIGger subsystem is used to set the controls and parameters
associated with triggering the data acquisitions. Other trigger-related
commands are found in the INITiate and ABORt subsystems.
External Trigger, Line and TV Trigger Delay
Value
:TRIGger[:SEQuence]:DELay <delay>
:TRIGger[:SEQuence]:DELay?
This command sets the amount of trigger delay when using the rear
panel external trigger input, the front panel input with TV trigger, or
the line trigger.
Factory Preset
1 µs
and *RST:
0.3 µs to 429 seconds
Range:
Default Unit:
seconds
External Trigger, Line and TV Trigger Delay
Enable
:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:DELay:STATe?
This command allows you to turn on or off a delay, during which the
analyzer will wait to begin a sweep after receiving an external trigger
signal, a front panel TV trigger, or a line trigger.
Factory Preset
and *RST:
Off
Default Unit:
seconds
Remarks:
Free-run activates the trigger condition that allows the
next sweep to start as soon as possible after the last
sweep. This function is not available when Gate is on.
Front Panel
Access:
Trig, Trig Delay On Off

External Trigger Slope

:TRIGger[:SEQuence]:EXTernal[1]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal[1]:SLOPe?
Chapter 5
Language Reference
TRIGger Subsystem
5-143