Declaration of Conformity for HOX2 Filter

We herewith inform you that the holmium oxide glass filter (type Hoya HY-1)
meets the following specification of absorbance maxima positions:
Agilent Technologies guarantees the traceability of the specified absorbance
maxima to a National Institute of Standards & Technology (NIST) holmium
oxide solution standard with a lot-to-lot tolerance of ±0.3nm.
The wavelength calibration filter built into the Agilent Technologies UV-visible
detectors
Table 54
Product No.
is made of this material and meets these specifications. It is, therefore,
suitable for wavelength calibration of these detectors within the specified
wavelength accuracy of the respective detector over its wavelength range.
310
(Part No. 79880-22711)
361.0 nm - 418.9 nm - 453.7 nm - 536.7 nm (spectral bandwidth: 2 nm).
79880A
79883A
79854A
1306A
G1315A/B
G1365A/B
G1600A
Engineering Manager
HP 1090 system
HP 1090 system
HP 1050 series
HP 1050 series
Agilent 1100 series
Agilent 1100 series
___March 20, 2000__
Date
1100 Series DAD and MWD Reference Manual
Wavelength Accuracy
± 1 nm
± 1 nm
± 1 nm
± 1 nm
± 1 nm
± 1 nm
± 1 nm
Quality Manager