Installing and Replacing the Cartridge

Installing and Replacing the
Cartridge
NOTE
Agilent CrossLab CS Electronic Leak Detector Operation Manual

Installing the cartridge

The CrossLab CS Mainframe is a platform that can accept the
ADM Flow Meter Cartridge P/N G6692A and Electronic Leak
Detector Cartridge P/N G6694A.
If you purchased the leak detector cartridge standalone,
update your existing mainframe to the latest firmware when
using your leak detector for the first time.
Remove the packaging on the Leak Detector Cartridge and
insert into the CrossLab CS Mainframe slot. Hand tighten the
two captive thumbscrews. The Leak Detector Cartridge comes
with an attached sniffer probe and tubing which is
not removable.
Figure 6. Leak Detector Cartridge being inserted into the CrossLab CS
Mainframe slot.
Captive screws