Optical System - Agilent Technologies InfinityLab LC Series User Manual

Diode array detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

1
Introduction

Optical System

Optical System

The optical system of the detector is shown in
Figure 4
1
2
3
4
5
6
7
The illumination source is a deuterium-arc-discharge lamp [1] for the ultraviolet
(UV) wavelength range. Its light is focused by a lamp mirror [2] onto the entrance
of the Max-light cartridge flow cell [3] with optofluidic waveguides. The light
leaves the Max-light cartridge flow cell at the other side and is focused by the fold
mirror [4] through the slit assembly [5] onto a holographic grating [6] light being
dispersed onto the diode array [7]. This allows simultaneous access to all
wavelength information.
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Optical System of the Detector
UV-lamp
Lamp mirror
Flow cell
Fold mirror
Programmable (G7117B) or Fixed (G7117A/G7117C) slit
Grating
Array
Figure 4
on page 14.
14

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents