4
Repairing the Detector

Cleaning or Replacing Coupling Lens Assembly

When required
Tools required
Parts required
Preparations for this procedure:
Turn off the lamp(s).
If the detector was in use, the lamp may be hot. If
so, wait five minutes until lamp cools off.
Remove the front cover.
Remove both lamps, see
page 102.
132
The coupling lens assembly is located between the tungsten and the
deuterium lamp. It may be replaced when light transmission is reduced and a
lamp exchange or window cleaning does not improve the light throughput.
Screwdriver POZI 1 PT3
hexagonal key 1.5 mm
A pointed pair of pliers
Coupling lens assembly G1103-68001
"Exchanging a Lamp"
on
1
Use the 1.5-mm hex key to loosen the setscrew that
secures the lens assembly.