Replacing The Holmium Oxide Filter Motor - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector

Replacing the Holmium Oxide Filter Motor

When required
Tools required
Parts required
If the filter motor was removed, the filter lever should not be reused. Always use a new
N O T E
filter lever to assure correct fit on the filter motor shaft.
Preparations for this procedure:
Turn off the lamp.
Switch off the detector, and disconnect the power cable.
Remove detector from stack and place it on working
bench.
Remove front cover, top cover and top foam section, see
"Removing the Top Cover and Foam"
Remove optical unit, see
(G1315A/G1365A)"
Optical Unit (G1315B/G1365B)"
162
If defective
Screwdriver POZI 1 PT3
hexagonal key 3 mm
Filter motor assembly G1315-68700 (includes filter lever G1315-45001 and
spring 1460-1510)
For others see
on page 140.
"Removing the Optical Unit
on page 154 or
"Removing the
on page 156.
"Holmium Oxide Filter"
on page 197
1 Place the optical unit on your bench.
1100 Series DAD and MWD Reference Manual